Back to Menu
🧈

PDFF Liver Fat Assistant

Enter the liver proton density fat fraction to get a one-line report with clinical significance.
%
No PDFF map? Estimate a signal fat fraction
Signal fat fraction = (IP − OP) ÷ (2 × IP), using the same ROI on both images, away from vessels and edges. Semiquantitative: iron (T2* decay at the later in-phase echo) and the multi-peak fat spectrum make it underestimate fat, T1 weighting makes it overestimate it, and two-point magnitude data cannot read above 50%. It is not PDFF, so it is not graded with the PDFF cutoffs; see section 7 below. Reeder et al., JMRI 2011.
Thresholds & reference
PDFFInterpretation
< 5%No hepatic steatosis (normal)
5 – <17.5%Mild steatosis (grade 1)
17.5 – <22%Moderate steatosis (grade 2)
≥ 22%Severe steatosis (grade 3)

Steatosis onset at PDFF >5% (EASL). Severity band edges (≈17.5%, ≈22%) follow Tang et al., Radiology 2013;267:422–431, and are approximate — histologic-grade cutoffs vary by study and are not universally standardized. A proper PDFF sequence corrects for T1 bias, T2*/iron, and fat spectral complexity; low SNR and motion still degrade accuracy. Interpret with clinical context.

The physics & pitfalls (fellow / attending level)

What PDFF actually measures. PDFF is the density of mobile fat protons divided by the density of mobile fat plus water protons. Once the confounders below are removed it is a property of the tissue, not of the scanner, which is why it agrees with MR spectroscopy (mean bias about −0.1 percentage points, R² 0.96) and why field strength, vendor and reconstruction have little effect on it. It is not the histology number. Pathologists grade the percentage of hepatocytes containing fat vacuoles; PDFF counts fat by proton density. The two scales are related but not equal, which is why the PDFF grade cutoffs (5, about 17.5, about 22%) sit well below the histologic ones (5, 33, 66%).

How fat and water are separated. The main fat peak (methylene) resonates about 3.4 ppm from water, roughly 220 Hz at 1.5 T and 440 Hz at 3 T, so fat and water drift in and out of phase every 4.6 ms at 1.5 T (2.3 ms at 3 T). A spoiled gradient echo samples that cycle with at least four, preferably six, closely spaced echoes, and a model of fat plus water plus decay is fitted to how the signal evolves. Fitting the complex data, phase as well as magnitude, tells the fit which component is dominant, so PDFF spans 0 to 100% instead of the 50% ceiling of two-point In-/Out-of-Phase imaging.

What the sequence corrects, and how. T1 bias: fat has a much shorter T1 than water and would be over-weighted, so the flip angle is kept low (for 3D, roughly 3–5° at 1.5 T and 2–4° at 3 T). T2* decay: modeled across the echoes as a single R2*, which also produces an R2* map as a by-product, a free iron screen on every PDFF exam (see the Iron tool). Fat spectral complexity: fat has about six peaks, not one, and a fixed six-peak model accounts for them. Noise bias: at low SNR magnitude data push the estimate away from the truth, which is why large voxels and adequate SNR matter. Eddy-current phase errors: handled in reconstruction, typically by hybrid magnitude and complex fitting.

Field strength is portable here. This is the opposite of the iron tool. R2* scales with field strength and its LIC calibration is 1.5 T only, but a confounder-corrected PDFF is the same number at 1.5 T and 3 T. No conversion is needed between exams on different magnets.

Fat-water swap. The fit has two near-equal solutions, fat-dominant and water-dominant, and field inhomogeneity can push it to the wrong one. A swapped region reads close to 100 minus the true value. It favors the dome, the lateral edges, tissue near metal or bowel gas, and sometimes a whole slice. Look at the PDFF map, not just the ROI number: a segment implausibly higher than its neighbors is a swap until the map says otherwise.

Iron and motion. With heavy iron the later echoes decay into noise and the fat estimate loses precision. The QIBA profile treats PDFF as unreliable above an R2* of about 300 s−1 at 1.5 T or 450 s−1 at 3 T (300 s−1 is an LIC of about 8 mg/g on the Iron tool's calibration). Check the R2* map before trusting a fat fraction in an iron-loaded liver. Breath-hold failure blurs the map and mixes in vessel and edge signal.

ROIs and change over time. The most reproducible method is the largest ROI that fits in each of the nine Couinaud segments, averaged, avoiding vessels, ducts, edges and artifact (Campo et al. 2017); QIBA asks for ROIs of at least 1 cm diameter and prefers 4 cm² or more. Precision: the repeatability coefficient is about 3 percentage points on the same scanner and about 4 across scanners or protocols (Yokoo et al. 2018), and QIBA treats an absolute change of 5 points or more as real with 95% confidence. In MASH treatment trials, a relative drop of 30% or more is associated with histologic response (51% vs 14%; OR about 7) and NASH resolution (Stine et al. 2021). Note the tension at low baseline: 30% of a 10% PDFF is 3 points, inside measurement noise, so compare on the same scanner and protocol and read small changes cautiously.

Imaging diagnosis of hepatic steatosis
The many ways to detect and quantify liver fat, from first-line ultrasound to the MRI reference standards. Expand any modality for principle, thresholds, performance, and pitfalls. MRI-PDFF (this calculator) and MR spectroscopy are the most accurate.
ModalityQuant?SensSpecMild?Monitor?Key threshold
B-mode USNo80–89%87–90%PoorNoLiver > renal cortex echogenicity
Hepato-renal indexSemi~83%~98%ModerateLimitedHRI ≥ 1.3
CAP (FibroScan)Semi80–92%~90%ModerateNo≥ 275 dB/m (AASLD)
US attenuation coeff.Yes~83%~77%ModerateUnder study≥ 0.66 dB/cm/MHz
Non-contrast CTSemi72% / 82%88% / 94%PoorNoLiver < 40 HU; L−S < −10 HU
Contrast-enhanced CTSemi66%90%PoorNoUnreliable (see note)
DECT fat fractionYes85%88%ModerateUnder studyFVF ≥ 5.3%
In- & Out-of-Phase MRI (2-point Dixon)Semi77–90%87–91%YesLimitedOut-of-Phase signal drop
MRI-PDFFYes77–100%87–100%YesYesPDFF ≥ 5–5.56%
MR spectroscopyReference80–98%80–98%YesYes (research)Fat fraction ≥ 5.56%
CT attenuation → estimated liver fat (Kodama 2007)
HU
—
Unenhanced: fat% = 100 · exp[−exp(−1.915 + 0.051 · HU)]. Portal venous: fat% = 100 · exp[−exp(−1.781 + 0.024 · HU)]. A continuous estimate of pathologic fat from mean liver attenuation. The portal-venous model is protocol/timing-dependent and less reliable — use DECT/VNC or MRI-PDFF for quantification. Kodama et al., AJR 2007;188:1307–1312.
1 · B-mode ultrasound (grayscale)First-line

Principle

Most widely used first-line test — low cost, no radiation, wide availability. Fat raises hepatic echogenicity.

Criteria

Hepato-renal index (HRI)

Semi-quantitative liver-to-renal-cortex echogenicity ratio. AUROC 0.91; optimal cutoff HRI ≥ 1.3.

Performance

Sensitivity 80–89%, specificity 87–90% for moderate-severe steatosis; sensitivity falls below 50% for mild steatosis.

Limitations

AASLD 2025 (Sterling) & 2023 (Rinella); AHA 2022 (Duell); Tilg et al., JAMA 2026.
2 · Controlled attenuation parameter (CAP / FibroScan)

Principle

Point-of-care measure of ultrasound attenuation through the liver, acquired with liver stiffness on vibration-controlled transient elastography.

Thresholds

Performance

Sensitivity ~80–90%, specificity ~90% for any steatosis.

Limitations

AASLD 2025 (Sterling) & 2023 (Rinella); Runge et al., Radiology 2018.
3 · Ultrasound attenuation coefficient (AC / ATT)Emerging

Principle

Newer quantitative tool built into conventional US machines; measures the attenuation coefficient of the beam as an add-on to B-mode.

Thresholds

Performance

AUROC 0.82–0.89 — comparable to or slightly better than CAP in some studies; scalable on routine machines.

Ferraioli & Soares Monteiro, World J Gastroenterol 2019; Collin et al. 2023; Gonçalves et al. 2026; AASLD 2025.
4 · Non-contrast CT (incl. liver–spleen)L−S valid

Principle

Opportunistic detection; fat lowers liver attenuation. Normal liver is ~8–10 HU denser than spleen.

Criteria (validated on non-contrast CT)

Performance

Sensitivity 72% / specificity 88% for any steatosis (≥5% fat); 82% / 94% for ≥ moderate (≥20–33% fat).

Continuous estimate

Beyond thresholds, mean liver attenuation maps to a pathologic fat % (see the Kodama estimator above): fat% = 100 · exp[−exp(−1.915 + 0.051 · HU)] on unenhanced CT.

Limitations

Haghshomar et al., Radiology 2024 (42 studies, 14,186 pts); Kodama et al., AJR 2007 (attenuation–fat equations); AASLD 2025; AHA 2022.
5 · Contrast-enhanced CT & VNCRatio unreliable

Why liver/spleen enhancement ratios fail here

After iodinated contrast, liver and spleen enhance on different time courses — early on the spleen shows heterogeneous arciform ("zebra") enhancement, and the liver–spleen relationship keeps shifting with phase, scan timing, contrast dose, and cardiac output. No single enhancement ratio or difference is reproducible across exams, so a value meaning steatosis at one timing means nothing at another. Absolute and liver/spleen-ratio thresholds are therefore validated only on non-contrast CT. On enhanced scans, use DECT fat fraction or a virtual non-contrast series instead of a ratio.

Performance (enhanced, absolute attenuation)

Sensitivity 66% / specificity 90% for any steatosis — lower than NCCT.

Portal-venous fat equation (historical)

Kodama derived a portal-venous model — fat% = 100 · exp[−exp(−1.781 + 0.024 · HU)] (in the estimator above) — but it is tied to that study's injection protocol and timing, so it does not generalize; treat any enhanced-CT fat number as a rough estimate, not a quantification.

Virtual non-contrast (VNC) from dual-source DECT

Approximates true unenhanced attenuation. VNC 46 HU threshold: 69% sensitivity / 95% specificity for ≥ moderate steatosis (vs. 100% / 42% at the standard 40 HU). Note VNC tends to overestimate true unenhanced HU.

Haghshomar et al., Radiology 2024; Catania et al., Eur J Radiol 2024; Kodama et al., AJR 2007.
6 · Dual-energy CT (DECT) fat fraction

Principle

Material decomposition generates fat-fraction maps and virtual monoenergetic images; can quantify fat even from contrast-enhanced acquisitions without an extra non-contrast dose.

Thresholds

Performance

Sensitivity 85% / specificity 88%; DECT-FVF correlates strongly with MRI-PDFF (Spearman r = 0.933).

Haghshomar et al., Radiology 2024; Hou et al., Eur J Radiol 2025.
7 · In- and Out-of-Phase (two-point Dixon)Not for quantification

Principle

Water and fat protons precess at different frequencies; signal loss on the Out-of-Phase image relative to the In-Phase image indicates intracellular lipid (Out-of-Phase = opposed-phase). Excellent for detecting and localizing fat.

The two-point fat-fraction equation

The commonly used estimate is FF = (SIIn − SIOut) / (2 × SIIn). It tracks fat at low–moderate levels in a clean liver, but it is a semiquantitative screen, not a true fat fraction — four confounders make it unreliable for quantification:

Spleen-normalized “fat %” does not rescue it

Normalizing liver to spleen at each echo (Borra 2009) assumes the two organs share the same T2* — the very assumption iron destroys — so it is arguably more fragile in the patients who most need accuracy.

Appropriate use

Treat the In- and Out-of-Phase result as qualitative-to-semiquantitative (steatosis present/absent, roughly graded) — not as a reportable fat fraction or PDFF.

Reliable alternative

Confounder-corrected multi-echo PDFF (section 8) explicitly models and removes T2*, the multi-peak fat spectrum, T1, and noise — the reference standard for quantification. Multi-point Dixon (3+ echoes) with T2* correction agrees closely with MRS (mean diff −0.5% ± 4.6%).

Reeder et al., JMRI 2011; Borra et al., Radiology 2009; Bashir et al. 2012; Hu et al., Radiology 2021 (multiplatform PDFF phantom); ACR Appropriateness Criteria 2023.
8 · MRI-PDFFThis tool · reference-grade

Principle

Multi-echo chemical-shift-encoded gradient echo with multipeak spectral fat modeling and T2* correction yields a quantitative fat-fraction map of the whole liver.

Thresholds

Performance

Sensitivity 77–92%, specificity 87–94% in most series, reaching 100% in the single study with multipeak and T2* correction noted above (hence 77–100% in the comparison table); AUROC 0.96–0.99 vs. histology; correlation with MRS r = 0.95–0.99.

Advantages

Limitations

AASLD 2025 (Sterling) & 2023 (Rinella); ACR Appropriateness Criteria 2023; Runge et al. 2018; Idilman et al. 2016.
9 · MR spectroscopy (¹H-MRS)Reference standard

Principle

Single-voxel proton MRS measures the lipid-to-water signal ratio in a defined liver volume — the non-invasive reference standard used to validate other methods.

Performance

Sensitivity and specificity 80–98%; excellent correlation with histology (r = 0.71–0.96).

Limitations

AASLD 2025 (accepted reference standard); ACR Appropriateness Criteria 2023; Idilman et al. 2016.
Guideline library

Educational summary compiled from the above; figures/percentages reflect the cited primary literature and may vary by population and technique.